HACH 145300

HACH Total Hardness Test Kit, Model 5-B

מאָדעל נומער: 145300

1. הקדמה

The HACH Total Hardness Test Kit, Model 5-B, provides a simple and economical method for determining water hardness through drop count titration. This kit is designed for field and laboratory use, offering reliable measurements of total hardness expressed as CaCO₃. The method utilizes a drop count titration with a range of 1-30 grains per gallon (gpg), with each drop of titrant equivalent to 1 gpg. The kit contains sufficient reagents for approximately 100 tests.

2. זיכערקייַט אינפֿאָרמאַציע

WARNING: This kit contains chemicals that may be harmful if misused. Carefully read the label on each container before use.

  • Always wear appropriate personal protective equipment, such as safety glasses, to protect eyes from chemical splashes.
  • Avoid direct contact of reagents with skin and clothing. In case of contact, rinse immediately with plenty of water.
  • Do not ingest any reagents. If ingested, seek immediate medical attention.
  • Wash hands thoroughly with soap and water after handling chemicals.
  • Dispose of all chemical waste according to local environmental regulations.

3. קיט קאַמפּאָונאַנץ

The HACH Total Hardness Test Kit, Model 5-B, includes the following items:

  • Titrant Solution, Hardness 3, 0.015 M EDTA (100 mL bottle with dropper)
  • UniVer® 3 Hardness Reagent (powder, 26.3 g bottle)
  • Measuring Tube (Cat. No. 438-00)
  • Square Mixing Bottle (Cat. No. 439-06)
  • Measuring Scoop (for UniVer® 3 Hardness Reagent)
Hach Total Hardness Test Kit, Model 5-B components
Image 1: The Hach Total Hardness Test Kit, Model 5-B, showing the titrant bottle, UniVer® 3 reagent, measuring tube, square mixing bottle, and measuring scoop.

4. סעטאַפּ

Before beginning the test, ensure all components are clean and readily accessible. Work in a well-lit area.

  1. Unpack all kit components and verify their presence against the list above.
  2. Rinse the measuring tube and square mixing bottle thoroughly with the water sample to be tested, then discard the rinse water. This prevents contamination from previous tests or tap water.

5. Operating Instructions: Drop Count Titration

Follow these steps to perform a total hardness test:

  1. קלייַבן Sampלאַ: Fill the measuring tube (Cat. No. 438-00) to the 25 mL mark with the water sampצו זיין טעסטעד.
  2. Transfer Sampלאַ: Pour the 25 mL water sample from the measuring tube into the square mixing bottle (Cat. No. 439-06).
  3. צולייגן רעאַגענט: Using the provided measuring scoop, add one level scoop of UniVer® 3 Hardness Reagent (Cat. No. 511-00) to the square mixing bottle. Cap the bottle and swirl gently to mix until the powder dissolves completely. The sample should turn a red or pink color if hardness is present. If the sample turns blue immediately, the water is very soft (less than 1 gpg).
  4. Titrate Sampלאַ: Hold the Titrant Solution bottle (0.015 M EDTA) vertically and add the solution drop by drop to the sample in the square mixing bottle. After each drop, cap the bottle and swirl to mix thoroughly.
  5. Observe Color Change: Continue adding drops and swirling until the color of the sample changes from red/pink to a clear blue. This is the endpoint of the titration.
  6. Count Drops: Record the total number of drops of Titrant Solution added to achieve the color change.

רעכענונג: Each drop of Titrant Solution corresponds to 1 grain per gallon (gpg) of total hardness. Therefore, the number of drops equals the total hardness of your water sample in gpg.

קאַנווערזשאַן: To convert grains per gallon (gpg) to milligrams per liter (mg/L) as CaCO₃, multiply the gpg value by 17.1. (1 gpg = 17.1 mg/L CaCO₃).

Person conducting a water test outdoors
Image 2: A person in protective gear conducting a water quality test in an outdoor industrial setting, demonstrating typical application environments for Hach products.

6. ינטערפּריטינג רעזולטאַטן

The total number of drops of Titrant Solution required to change the sample color from red/pink to blue directly indicates the water hardness in grains per gallon (gpg). For example, if 10 drops were required, the water hardness is 10 gpg.

Understanding your water hardness is crucial for various applications, including water treatment, appliance maintenance, and industrial processes.

Scientist in a lab coat using a Hach device
Image 3: A scientist in a lab coat and safety glasses operating a Hach water analysis device in a laboratory setting, highlighting the professional use of Hach equipment.

7. וישאַלט

Proper maintenance ensures the longevity and accuracy of your test kit.

  • רייניקונג: After each use, thoroughly rinse the measuring tube, square mixing bottle, and scoop with distilled or deionized water. Allow them to air dry completely before storage.
  • סטאָרידזש: Store all reagents in their original containers, tightly capped, in a cool, dark place away from direct sunlight and extreme temperatures. Ensure the Titrant Solution bottle is stored upright.
  • רעאַגענט עקספּיראַציע: Check the expiration dates on the reagent bottles. Expired reagents may lead to inaccurate results. Replace reagents as needed.

8. טראָובלעשאָאָטינג

If you encounter issues during testing, consider the following common problems and solutions:

  • No Color Change to Red/Pink: אויב די sample does not turn red/pink after adding UniVer® 3 Hardness Reagent, the water may be extremely soft (less than 1 gpg) or the reagent may be expired. Try a known hard water sample to verify reagent activity.
  • ומפּינקטלעך רעזולטאטן:
    • Ensure all glassware is clean and free from contamination.
    • Verify that reagents are not expired.
    • Ensure accurate measurement of the water sample (25 mL).
    • Add titrant drops consistently and swirl thoroughly after each drop.
  • Difficulty Seeing Color Change: Perform the test under good lighting conditions. A white background can help make the color change more apparent.

9. ספּעסאַפאַקיישאַנז

באַשרייַבונגדעטאַל
מאָדעל נומער145300
פּאַראַמעטערHardness, Total - As CaCO₃
אופֿן נאָמעןDrop Count Titration / EDTA-bulk powder
קייט1-30 gpg (grains per gallon)
Approximate Tests100
פּעקל דימענשאַנז5.04 X 2.68 X 2.6 אינטשעס
פּונקט וואָג8.96 אונסעס
פאַבריקאַנטHACH פירמע

10. וואָראַנטי און שטיצן

HACH products are manufactured to high-quality standards. For specific warranty information, please refer to the documentation included with your purchase or visit the official HACH website. For technical support, product inquiries, or to purchase replacement reagents, please contact HACH customer service or visit the HACH store online.

HACH Store: https://www.amazon.com/stores/Hach/page/F31EE231-E533-4E96-B7EB-49EC9C17D676

פֿאַרבונדענע דאָקומענטן - 145300

Preview HACH Hardness (Total and Calcium) Test Kit HA-4P MG/L (145701) - Instructions and Specifications
Detailed instructions and specifications for the HACH Hardness (Total and Calcium) Test Kit, Model HA-4P MG/L (145701). Learn how to perform total and calcium hardness tests, understand interferences, and order replacement items.
Preview HACH EZ4041 Total Hardness Analyser - Water Quality Monitoring
Discover the HACH EZ4041 Total Hardness Analyser, an online, automatic colorimetric titrator designed for industrial and environmental water monitoring. Features include a wide measuring range, smart automatic functions, flexible titration methods, and multiple stream analysis capabilities.
Preview Hach Methods Quick Reference Guide: Chemical Analysis Instruments and Reagents
A comprehensive guide to Hach's chemical analysis methods, detailing test parameters, EPA approvals, instrument compatibility, and product numbers for water quality testing.
Preview How to Reprogram Hach DR-3000 for Accu-TEST COD Vials
Instructions and services for reprogramming the Hach DR-3000 spectrophotometer or colorimeter to use Bioscience, Inc.'s accu-TEST COD vials. Includes support and reprogramming service details from Bioscience, Inc.
Preview Hach NitraVer 5 Nitrate High Range Method 8039: Analysis Guide
Comprehensive guide to performing high-range nitrate analysis (0-30.0 mg/L NO3--N) using Hach's NitraVer 5 reagent with Powder Pillows or AccuVac Ampuls (Method 8039).
Preview Hach Chlorine Test Kit Instructions: Quick Guide for Water Quality Testing
This quick guide provides instructions for using the Hach Chlorine Test Kit to measure free or total chlorine levels in water. It includes preparation steps, a detailed procedure, and important notices for accurate water quality testing.